The Ultrafast Systems imaging monochromator is optimized to work with our time-resolved spectrometers. This monochromator features high spectral resolution, as well as an astigmatism-corrected design, optimal for multipixel array detectors. With its dual exit ports, this instrument can be used as a spectrograph as well as a monochromator.
The monochromator’s focusing matches the design of the main optical bench to maximize the overall spectrometer’s performance.
The monochromator features a 4-grating turret with kinematic mounts for optimal wavelength coverage, optical throughput, and spectral resolution.
The UV-enhanced Al coatings on the monochromator mirrors ensure high efficiency of fluorescence and the upconverted signal collection as well as excellent performance in Mid-IR transient absorption experiments.
Focal Length, mm
300
Aperture
f/6
Spectral Range
200 nm to 20 µm
Grating Size, mm
50 x 50
Number of Gratings
up to 4
Number of Entrance Ports
1
Entrance Slit Width (adjustable), mm
0.01 - 6
Number of Exit Ports
2
Flat Field Size, mm
30 x 7
Magnification
1x
Dimensions, mm
L355 x W305 x H266
50 l/mm
64 nm/mm
300 l/mm
11 nm/mm
1200 l/mm
2.7 nm/mm
UV-VIS-NIR
IR
Spectral Range
550-2000 nm
200-700 nm
200-700 nm
6-16 µm
3-10 µm
2-7 µm
Groove Density, g/mm
300
600
1200
50
75
150
Blaze Wavelength
760 nm
300 nm
300 nm
9 µm
4.65 µm
3 µm
Linear Dispersion, nm/mm
10.3
5
2.3
63.6
42.4
21.3
Mechanical Resolution, nm
0.62
0.3
0.14
3.8
2.5
1.3
Minimum Step Size, nm
0.29
0.14
0.07
1.7
1.2
0.57